Non-Rechargeable Replacement battery for the Physio-Control LifePak 500 Defibrillator . Replaces 3005380-026, 11141-000013, 3208231-000, 11141-000147 and 3200390.
Chemistry:Sealed Li-So2Rating: 12V 7500mAhTypical Capacity: 14 hrs of "ON" time / 312 Full DischargesMinimum Capcity: 230 Full Discharges
Shelf Life: A new lithium battery pack has a shelf life of five (5) years stored at proper temperatures. Although a lithium battery may be stored for up to five (5) years, it self-discharges over time. Therefore, when the battery is eventually placed in the AED, it's useful life will be reduced depending on how long it "sat on the shelf"
Recommended Replace Interval: Three (3) years, depending upon usage.
Self Testing: The non-rechargeable lithium battery pack requires less maintenance then the rechargeable SLA battery pack since it never requires charging. With the lithium battery pack installed, the LIFEPAK 500 AED automatically tests it as part of the Daily Auto test. The AED also performs the battery test the first time the AED is turned on after a new battery has been installed.